ASTM-F2299 2003
$40.63
F2299-03 Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
Published By | Publication Date | Number of Pages |
ASTM | 2003 | 8 |
ASTM F2299-03
Historical Standard: Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
ASTM F2299
Scope
1.1 This test method establishes procedures for measuring the initial particle filtration efficiency of materials used in medical facemasks using monodispersed aerosols.
1.1.1 This test method utilizes light scattering particle counting in the size range of 0.1 to μ5.0 m and airflow test velocities of 0.5 to 25 cm/s.
1.2 The test procedure measures filtration efficiency by comparing the particle count in the feed stream (upstream) to that in the filtrate (downstream).
1.3 The values stated in SI units or in other units shall be regarded separately as standard. The values stated in each system must be used independently of the other, without combining values in any way.
1.4 The following precautionary caveat pertains only to the test methods portion, Section 10, of this specification. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Keywords
filtration; medical face mask; particle filtration efficiency; particles; particulate
ICS Code
ICS Number Code 11.120.20 (Wound dressings. Compresses)
DOI: 10.1520/F2299-03
PDF Catalog
PDF Pages | PDF Title |
---|---|
1 | Scope Referenced Documents Terminology |
2 | Summary of Test Method Significance and Use Apparatus System Preparation and Control |
3 | FIG. 1 |
4 | FIG. 2 FIG. 3 |
5 | Number of Downstream/Upstream Sampling Intervals TABLE 1 |
6 | Material Specimen Selection and Conditioning Test Procedure Calculations |
7 | Report Precision and Bias TABLE 2 TABLE 3 TABLE 4 TABLE 5 TABLE 6 |
8 | Keywords |