BS EN IEC 60601-2-31:2020
$198.66
Medical electrical equipment – Particular requirements for the basic safety and essential performance of external cardiac pacemakers with internal power source
Published By | Publication Date | Number of Pages |
BSI | 2020 | 60 |
IEC 60601-2-31:2020 is available as IEC 60601-2-31:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 60601-2-31:2020 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of EXTERNAL PACEMAKERS powered by an INTERNAL ELECTRICAL POWER SOURCE, hereafter referred to as ME EQUIPMENT. This document applies to PATIENT CABLES as defined in 201.3.209, but does not apply to LEADS as defined in 201.3.206. HAZARDS inherent in the intended physiological function of ME EQUIPMENT within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. This document does not apply to the implantable parts of ACTIVE IMPLANTABLE MEDICAL DEVICES covered by ISO 14708-1. This document does not apply to EXTERNAL PACEMAKERS which can be connected directly or indirectly to a SUPPLY MAINS. This document does not apply to transthoracic and oesophageal pacing ME EQUIPMENT and antitachycardia ME EQUIPMENT. IEC 60601-2-31:2020 cancels and replaces the second edition published in 2008 and Amendment 1:2011. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) The requirement for testing for energy reduction has been removed; b) The test for exposure to external defibrillation has been completely revised; c) The exclusion for testing ESD immunity only with respect to air discharges has been removed; d) Alignment with the latest edition of ISO 14708-2 for pacemakers, as well as the associated EMC standard ISO 14117; e) Additional rationale for all changes.
PDF Catalog
PDF Pages | PDF Title |
---|---|
2 | undefined |
14 | Table 201.101 – Distributed essential performance requirements |
16 | Table 201.102 – Dual chamber connector terminal marking |
20 | Figure 201.101 – Test waveform V test implemented by example RCL circuit using C = 120 μF, L = 25 μH, RL + R = 1 Ω |
21 | Figure 201.102 – Example circuit of defibrillation test voltage generator for generating a decaying exponential waveform |
22 | Figure 201.103 – Test setup for a single chamber external cardiac pacemaker Figure 201.104 – Test setup for a dual chamber external cardiac pacemaker |
23 | Figure 201.105 – Test setup for a triple chamber external cardiac pacemaker, e.g. bi-ventricular external cardiac pacemaker Figure 201.106 – Timing sequence |
25 | Figure 201.107 – Measuring circuit for the patient auxiliary current for me equipment with an internal electrical power source |
27 | Table 201.103 – me equipment parameters |
28 | Figure 201.108 – Measuring circuit for the maximum tracking rate |
29 | Figure 201.109 – Initial oscilloscope display when measuring maximum tracking rate |
32 | Table 202.101 – Static discharge requirements |
33 | Annex I Identification of immunity pass/fail criteria |
34 | Annex AA (informative)Particular guidance and rationale |
35 | Table AA.1 – External pacemaker hazard inventory |
41 | Figure AA.1 – Simple model of a single chamber external pacemaker during defibrillation |
43 | Figure AA.2 – First proposal for a defib-protection testof single chamber external pacemaker |
44 | Figure AA.3 – Circuit for a defibrillation test generator for defibrillation test according to conditions during open heart surgery |
45 | Figure AA.4 – Defibrillation pulse generated bythe defibrillation test generator from Figure AA.3 |
46 | Table AA.2 – Pulse energies calculated for C = 120 µF ± 5 % |
47 | Table AA.3 – Pulse energies calculated for C = 122 µF ± 5 % |
48 | Table AA.4 – Pulse energies calculated for C = 126,32 µF ± 5 % |
49 | Figure AA.5 – Rise times of a defibrillation pulse according to the circuit proposed in Figure AA.3 |