11.040.10 – Anaesthetic, respiratory and reanimation equipment – PDF Standards Store ?u= Wed, 06 Nov 2024 02:54:17 +0000 en-US hourly 1 https://wordpress.org/?v=6.7.1 ?u=/wp-content/uploads/2024/11/cropped-icon-150x150.png 11.040.10 – Anaesthetic, respiratory and reanimation equipment – PDF Standards Store ?u= 32 32 JIS T 7209:2007 ?u=/product/publishers/jis/jis-t-72092007/ Wed, 06 Nov 2024 02:54:17 +0000 Oxygen concentrators for medical use - Safety requirements
Published By Publication Date Number of Pages
JIS 2007-01-25 31
]]>
Oxygen concentrators for medical use - Safety requirements
Published By Publication Date Number of Pages
JIS 2007-01-25 31
]]>
JIS T 7205:1989 ?u=/product/publishers/jis/jis-t-72051989/ Wed, 06 Nov 2024 02:54:16 +0000 Manual (Operator-powered) Resuscitators
Published By Publication Date Number of Pages
JIS 1989-10-25 17
]]>
Manual (Operator-powered) Resuscitators
Published By Publication Date Number of Pages
JIS 1989-10-25 17
]]>
JIS T 7206:1989 ?u=/product/publishers/jis/jis-t-72061989/ Wed, 06 Nov 2024 02:54:16 +0000 Gas-powered Resuscitators
Published By Publication Date Number of Pages
JIS 1989-10-25 21
]]>
Gas-powered Resuscitators
Published By Publication Date Number of Pages
JIS 1989-10-25 21
]]>
JIS T 7204:1989 ?u=/product/publishers/jis/jis-t-72041989/ Wed, 06 Nov 2024 02:54:15 +0000 Lung Ventilators for Medical Use
Published By Publication Date Number of Pages
JIS 1989-06-01 31
]]>
Lung Ventilators for Medical Use
Published By Publication Date Number of Pages
JIS 1989-06-01 31
]]>
ISO/TS 81060-5:2020 ?u=/product/publishers/iso/iso-ts-81060-52020/ Wed, 06 Nov 2024 01:35:30 +0000 Non-invasive sphygmomanometers — Part 5: Requirements for the repeatability and reproducibility of NIBP simulators for testing of automated non-invasive sphygmomanometers
Published By Publication Date Number of Pages
ISO 2020-02 18
]]>
This document specifies requirements for the repeatability and reproducibility of non-invasive blood pressure (NIBP) simulators intended to test automated sphygmomanometers utilizing the oscillometric non-continuous method only.

In addition, the pulse rate set on the NIBP simulator is tested.

This document is not intended to relate the signals, generated by the NIBP simulator, to the oscillometric signal recorded in a cuff attached to a human. It does not intend to test the interaction between the NIBP simulator and the tested automated sphygmomanometer (e.g. the agreement of the set values of the NIBP simulator and the displayed values of the tested automated sphygmomanometer or the properties of the cuff and tubing, such as design or elastic properties).

NOTE 1 These parameters can be tested separately in a clinical investigation or by using different special test setups.

This document does not check whether or not the NIBP simulator is able to test the accuracy of the absolute blood pressure value of oscillometric automated sphygmomanometers.

NOTE 2 Usually this is tested by a clinical investigation according ISO 81060-2 or other protocols.

This document is applicable to NIBP simulators testing automated sphygmomanometers for adults, children and neonates at the upper arm, thigh etc. and automated sphygmomanometers measuring at the wrist.

]]>
ISO/TS 18835:2004 ?u=/product/publishers/iso/iso-ts-188352004/ Wed, 06 Nov 2024 01:29:54 +0000 Inhalational anaesthesia systems — Draw-over vaporizers and associated equipment
Published By Publication Date Number of Pages
ISO 2004-07 26
]]>
ISO/TS 18835:2004 specifies safety and performance requirements for draw-over vaporizers and associated equipment to provide draw-over anaesthetic systems for patients weighing greater than 15 kg using both non-flammable and, in places where regulations permit their use, flammable anaesthetic agents.

No requirements for monitoring the equipment are given in ISO/TS 18835:2004.

]]>
ISO/TS 16628:2003 ?u=/product/publishers/iso/iso-ts-166282003/ Wed, 06 Nov 2024 01:27:51 +0000 Tracheobronchial tubes — Recommendations for size designation and labelling
Published By Publication Date Number of Pages
ISO 2003-08 14
]]>
ISO/TS 16628:2003 provides recommendations for the size designation and labelling of tracheobronchial tubes, including colour coding of the bronchial cuff and its associated pilot balloon. Tracheobronchial tubes that include bronchus blockers are excluded from the scope of this Technical Specification.

]]>
ISO/TR 21954:2018 ?u=/product/publishers/iso/iso-tr-219542018/ Wed, 06 Nov 2024 01:15:33 +0000 Guidance on the selection of the appropriate means of ventilation based on the intended patient, use environment, and operator
Published By Publication Date Number of Pages
ISO 2018-07 48
]]>
This document considers and identifies criteria about the intended patient, intended use environment, and intended operator across the spectrum of the types of ventilation-related equipment as listed below:

— gas-powered resuscitator as specified in ISO 10651-5[1] [1];

— operator-powered resuscitator as specified in ISO 10651-4[2];

— ventilator for critical care as specified in ISO 80601-2-12[3] [2];

— ventilator for emergency medical services environment as specified in ISO 80601-2-84[4] [3], the future replacement for ISO 10651-3[5];

NOTE 1 ISO 80601-2-84 updates the content of ISO 10651-3 and harmonizes it with IEC 60601-1:2005+AMD1:2012[6] and IEC 60601-1-12:2014[7].

— ventilator for ventilatory impairment in the home healthcare environment as specified in ISO 80601-2-79[8];

— ventilator for ventilatory insufficiency in the home healthcare environment as specified in ISO 80601-2-80[9];

— ventilator for ventilator-dependent patients in the home healthcare environment as specified in ISO 80601-2-72[10];

— sleep apnoea breathing therapy equipment as specified in ISO 80601-2-70[11].

NOTE 2 Sleep apnoea breathing therapy equipment is not considered to be an artificial ventilator. It is included in this discussion to highlight the differences, which indicate why sleep apnoea breathing therapy equipment is not considered a ventilator.

This document is intended to provide guidance that can assist manufacturers, authorities having jurisdiction and users in the development, selection and application of different types of ventilatory equipment based on the intended patient, intended use environment and intended operator.

[1] Numbers in square brackets refer to the Bibliography.

[2] Under preparation. Stage at the time of publication: ISO/FDIS 80601-2-12:2018.

[3] Under preparation. Stage at the time of publication: ISO/DIS 80601-2-84:2018.

]]>
ISO 9919:2005 ?u=/product/publishers/iso/iso-99192005/ Wed, 06 Nov 2024 00:31:27 +0000 Medical electrical equipment — Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use
Published By Publication Date Number of Pages
ISO 2005-03 92
]]>
For the purposes of ISO 9919:2005, the Scope given in IEC 60601-1:1988 (Clause 1) applies, except as follows.

Amendment (add at the end of 1.1):

ISO 9919:2005 specifies particular requirements for the basic safety and essential performance of pulse oximeter equipment intended for use on humans. This includes any part necessary for normal use, e.g. the pulse oximeter monitor, pulse oximeter probe, probe cable extender.

These requirements also apply to pulse oximeter equipment, including pulse oximeter monitors, pulse oximeter probes and probe cable extenders, that has been reprocessed.

The intended use of pulse oximeter equipment includes, but is not limited to, the estimation of arterial oxygen haemoglobin saturation and pulse rate on patients in healthcare institutions as well as on patients in home care.

ISO 9919:2005 is not applicable to pulse oximeter equipment intended for use in laboratory research applications nor to oximeters that requires a blood sample from the patient.

ISO 9919:2005 is not applicable to pulse oximeter equipment solely intended for fetal use.

ISO 9919:2005 is not applicable to remote or slave (secondary) devices that display SpO2 values that are located outside of the patient environment.

The requirements of ISO 9919:2005 which replace or modify requirements of IEC 60601-1:1988 and its Amendments 1 (1991) and 2 (1995) are intended to take precedence over the corresponding general requirements.

]]>
ISO 9918:1993 ?u=/product/publishers/iso/iso-99181993/ Wed, 06 Nov 2024 00:31:26 +0000 Capnometers for use with humans — Requirements
Published By Publication Date Number of Pages
ISO 1993-02 32
]]>
Specifies requirements for the safety of capnometers and refers to and is applicable in conjunction with IEC 601-1 (Safety of medical electrical equipment, General safety requirements). Applies to capnometers used with adults, children and neonates. Does not apply to devices intended for use as transcutaneous monitors or intended for use in laboratory research applications.

]]>