{"id":293955,"date":"2024-10-19T19:58:49","date_gmt":"2024-10-19T19:58:49","guid":{"rendered":"https:\/\/pdfstandards.shop\/product\/uncategorized\/bs-en-iso-4064-22017\/"},"modified":"2024-10-25T17:11:05","modified_gmt":"2024-10-25T17:11:05","slug":"bs-en-iso-4064-22017","status":"publish","type":"product","link":"https:\/\/pdfstandards.shop\/product\/publishers\/bsi\/bs-en-iso-4064-22017\/","title":{"rendered":"BS EN ISO 4064-2:2017"},"content":{"rendered":"
This part of ISO 4064|OIML R 49 is applicable to the type evaluation and initial verification testing of water meters for cold potable water and hot water as defined in ISO 4064-1:2014|OIML R 49-1:2013. OIML Certificates of Conformity can be issued for water meters under the scope of the OIML Certificate System, provided that this part of ISO 4064|OIML R 49, ISO 4064-1:2014|OIML R 49-1:2013 and ISO 4064-3:2014|OIML R 49-3:2013 are used in accordance with the rules of the System.<\/p>\n
This part of ISO 4064|OIML R 49 sets out details of the test programme, principles, equipment and procedures to be used for the type evaluation, and initial verification of a meter type.<\/p>\n
The provisions of this part of ISO 4064|OIML R 49 also apply to ancillary devices, if required by national regulations.<\/p>\n
The provisions include requirements for testing the complete water meter and for testing the measurement transducer (including the flow or volume sensor) and the calculator (including the indicating device) of a water meter as separate units.<\/p>\n
PDF Pages<\/th>\n | PDF Title<\/th>\n<\/tr>\n | ||||||
---|---|---|---|---|---|---|---|
2<\/td>\n | National foreword <\/td>\n<\/tr>\n | ||||||
5<\/td>\n | Annex ZA(informative)Relationship between this European Standard and the essential requirements of Directive 2014\/32\/EU aimed to be covered <\/td>\n<\/tr>\n | ||||||
13<\/td>\n | Foreword <\/td>\n<\/tr>\n | ||||||
14<\/td>\n | Section sec_1 Section sec_2 1 Scope 2 Normative references <\/td>\n<\/tr>\n | ||||||
15<\/td>\n | Section sec_3 Section sec_4 3 Terms and definitions 4 Reference conditions <\/td>\n<\/tr>\n | ||||||
16<\/td>\n | Section sec_5 Section sec_6 Section sec_6.1 Section sec_6.2 Section sec_6.3 5 Symbols, units and equations 6 External examination 6.1 General 6.2 Object of the examination 6.3 Preparation <\/td>\n<\/tr>\n | ||||||
17<\/td>\n | Section sec_6.4 Section sec_6.4.1 Section sec_6.4.2 Section sec_6.4.3 Section sec_6.4.3.1 Section sec_6.4.3.2 Section sec_6.4.3.3 6.4 Examination procedures <\/td>\n<\/tr>\n | ||||||
18<\/td>\n | Section sec_6.4.3.4 Section sec_6.4.3.5 Section sec_6.4.3.5.1 Section sec_6.4.3.5.2 <\/td>\n<\/tr>\n | ||||||
19<\/td>\n | Section sec_6.4.3.5.3 Section sec_6.4.3.6 Section sec_6.4.3.6.1 Section sec_6.4.3.6.2 Section sec_6.4.3.6.2.1 <\/td>\n<\/tr>\n | ||||||
20<\/td>\n | Section sec_6.4.3.6.2.2 Section sec_6.4.3.6.2.3 <\/td>\n<\/tr>\n | ||||||
21<\/td>\n | Section sec_6.4.4 Section sec_7 Section sec_7.1 Section sec_7.2 Section sec_7.2.1 Section sec_7.2.2 Section sec_7.2.2.1 7 Performance tests for all water meters 7.1 General 7.2 Required conditions for all tests <\/td>\n<\/tr>\n | ||||||
22<\/td>\n | Section sec_7.2.2.2 Section sec_7.2.2.3 Section sec_7.3 Section sec_7.3.1 Section sec_7.3.2 Section sec_7.3.3 Section sec_7.3.3.1 Section sec_7.3.3.2 7.3 Static pressure test (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 4.2.10) <\/td>\n<\/tr>\n | ||||||
23<\/td>\n | Section sec_7.3.4 Section sec_7.4 Section sec_7.4.1 Section sec_7.4.2 Section sec_7.4.2.1 7.4 Determination of intrinsic errors (of indication) (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 7.2.3) <\/td>\n<\/tr>\n | ||||||
24<\/td>\n | Section sec_7.4.2.2 Section sec_7.4.2.2.1 Section sec_7.4.2.2.2 Section sec_7.4.2.2.3 <\/td>\n<\/tr>\n | ||||||
25<\/td>\n | Section sec_7.4.2.2.4 Section sec_7.4.2.2.4.1 Section sec_7.4.2.2.4.2 Section sec_7.4.2.2.4.3 Section sec_7.4.2.2.4.4 Section sec_7.4.2.2.4.5 <\/td>\n<\/tr>\n | ||||||
26<\/td>\n | Section sec_7.4.2.2.4.6 Section sec_7.4.2.2.5 Section sec_7.4.2.2.5.1 Section sec_7.4.2.2.5.2 Section sec_7.4.2.2.5.3 <\/td>\n<\/tr>\n | ||||||
27<\/td>\n | Section sec_7.4.2.2.6 Section sec_7.4.2.2.6.1 Section sec_7.4.2.2.6.2 Section sec_7.4.2.2.7 Section sec_7.4.2.2.7.1 Section sec_7.4.2.2.7.2 Section sec_7.4.2.2.7.3 <\/td>\n<\/tr>\n | ||||||
28<\/td>\n | Section sec_7.4.2.2.7.4 Section sec_7.4.2.2.7.5 <\/td>\n<\/tr>\n | ||||||
29<\/td>\n | Section sec_7.4.3 Section sec_7.4.3.1 Section sec_7.4.3.2 Section sec_7.4.4 <\/td>\n<\/tr>\n | ||||||
30<\/td>\n | Section sec_7.4.5 Section sec_7.4.6 Section sec_7.4.6.1 Section sec_7.4.6.2 Section sec_7.4.6.3 <\/td>\n<\/tr>\n | ||||||
31<\/td>\n | Section sec_7.4.6.4 Section sec_7.5 Section sec_7.5.1 Section sec_7.5.2 Section sec_7.5.3 Section sec_7.5.4 Section sec_7.6 Section sec_7.6.1 Section sec_7.6.2 Section sec_7.6.3 7.5 Water temperature test (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 4.2.8) 7.6 Overload water temperature test (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 7.2.5) <\/td>\n<\/tr>\n | ||||||
32<\/td>\n | Section sec_7.6.4 Section sec_7.7 Section sec_7.7.1 Section sec_7.7.2 Section sec_7.7.3 Section sec_7.7.4 Section sec_7.8 Section sec_7.8.1 7.7 Water pressure test (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 4.2.8) 7.8 Reverse flow test (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 4.2.7) <\/td>\n<\/tr>\n | ||||||
33<\/td>\n | Section sec_7.8.2 Section sec_7.8.3 Section sec_7.8.3.1 Section sec_7.8.3.2 Section sec_7.8.3.3 <\/td>\n<\/tr>\n | ||||||
34<\/td>\n | Section sec_7.8.4 Section sec_7.9 Section sec_7.9.1 Section sec_7.9.2 7.9 Pressure loss test (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 6.5) <\/td>\n<\/tr>\n | ||||||
35<\/td>\n | Table tab_b Figure fig_1 Section sec_7.9.3 Section sec_7.9.3.1 Section sec_7.9.3.2 <\/td>\n<\/tr>\n | ||||||
37<\/td>\n | Table tab_c Figure fig_2 Section sec_7.9.4 Section sec_7.9.5 <\/td>\n<\/tr>\n | ||||||
38<\/td>\n | Section sec_7.10 Section sec_7.10.1 Section sec_7.10.2 Section sec_7.10.3 Section sec_7.10.4 7.10 Flow disturbance tests (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 6.3.4) <\/td>\n<\/tr>\n | ||||||
39<\/td>\n | Section sec_7.11 Section sec_7.11.1 Section sec_7.11.2 Section sec_7.11.2.1 Section sec_7.11.2.2 Section sec_7.11.2.2.1 Section sec_7.11.2.2.2 Section sec_7.11.2.2.3 7.11 Durability tests (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 7.2.6) <\/td>\n<\/tr>\n | ||||||
40<\/td>\n | Section sec_7.11.2.2.4 Section sec_7.11.2.3 Section sec_7.11.2.3.1 Section sec_7.11.2.3.2 Section sec_7.11.2.3.3 Section sec_7.11.2.3.4 <\/td>\n<\/tr>\n | ||||||
41<\/td>\n | Section sec_7.11.2.3.5 Section sec_7.11.2.3.6 Table tab_1 <\/td>\n<\/tr>\n | ||||||
42<\/td>\n | Section sec_7.11.2.4 Section sec_7.11.2.4.1 Section sec_7.11.2.4.2 Section sec_7.11.3 Section sec_7.11.3.1 Section sec_7.11.3.2 Section sec_7.11.3.2.1 Section sec_7.11.3.2.2 <\/td>\n<\/tr>\n | ||||||
43<\/td>\n | Section sec_7.11.3.2.3 Section sec_7.11.3.3 Section sec_7.11.3.3.1 Section sec_7.11.3.3.2 Section sec_7.11.3.3.3 Section sec_7.11.3.3.4 <\/td>\n<\/tr>\n | ||||||
44<\/td>\n | Section sec_7.11.3.3.5 Section sec_7.11.3.4 Section sec_7.11.3.4.1 Section sec_7.11.3.4.2 Section sec_7.12 Section sec_7.13 Section sec_7.13.1 7.12 Magnetic field testing 7.13 Tests on ancillary devices of a water meter <\/td>\n<\/tr>\n | ||||||
45<\/td>\n | Section sec_7.13.2 Section sec_7.13.3 Section sec_7.13.4 Section sec_7.14 Section sec_8 Section sec_8.1 Section sec_8.1.1 7.14 Environmental testing 8 Performance tests related to influence factors and disturbances 8.1 General requirements (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.1) <\/td>\n<\/tr>\n | ||||||
46<\/td>\n | Section sec_8.1.2 Section sec_8.1.3 Section sec_8.1.4 Section sec_8.1.5 Section sec_8.1.6 <\/td>\n<\/tr>\n | ||||||
47<\/td>\n | Section sec_8.1.7 Section sec_8.1.8 Section sec_8.1.8.1 <\/td>\n<\/tr>\n | ||||||
48<\/td>\n | Section sec_8.1.8.2 Section sec_8.1.8.3 Section sec_8.1.8.4 <\/td>\n<\/tr>\n | ||||||
49<\/td>\n | Section sec_8.1.8.5 Section sec_8.2 Section sec_8.2.1 Section sec_8.2.2 Section sec_8.2.3 8.2 Dry heat (non-condensing) (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
50<\/td>\n | Section sec_8.2.4 Section sec_8.3 Section sec_8.3.1 Section sec_8.3.2 Section sec_8.3.3 Section sec_8.3.4 8.3 Cold (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
51<\/td>\n | Section sec_8.4 Section sec_8.4.1 Section sec_8.4.2 Section sec_8.4.3 8.4 Damp heat, cyclic (condensing) (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
52<\/td>\n | Section sec_8.4.4 Section sec_8.5 Section sec_8.5.1 Section sec_8.5.2 Section sec_8.5.2.1 Section sec_8.5.2.2 Section sec_8.5.2.3 8.5 Power supply variation (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
53<\/td>\n | Figure fig_3 Section sec_8.5.2.4 <\/td>\n<\/tr>\n | ||||||
54<\/td>\n | Section sec_8.5.3 Section sec_8.5.3.1 Section sec_8.5.3.2 Section sec_8.5.3.3 Section sec_8.5.3.4 Section sec_8.5.4 Section sec_8.5.4.1 <\/td>\n<\/tr>\n | ||||||
55<\/td>\n | Section sec_8.5.4.2 Section sec_8.5.4.3 Section sec_8.6 Section sec_8.6.1 Section sec_8.6.2 Section sec_8.6.3 8.6 Vibration (random) (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
56<\/td>\n | Section sec_8.6.4 Section sec_8.7 Section sec_8.7.1 Section sec_8.7.2 Section sec_8.7.3 8.7 Mechanical shock (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
57<\/td>\n | Section sec_8.7.4 Section sec_8.8 Section sec_8.8.1 Section sec_8.8.2 Section sec_8.8.3 8.8 AC mains voltage dips, short interruptions and voltage variations (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
58<\/td>\n | Table tab_2 Table tab_3 Section sec_8.8.4 Section sec_8.9 Section sec_8.9.1 8.9 Bursts on signal lines (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
59<\/td>\n | Section sec_8.9.2 Section sec_8.9.3 Section sec_8.9.4 <\/td>\n<\/tr>\n | ||||||
60<\/td>\n | Section sec_8.10 Section sec_8.10.1 Section sec_8.10.2 Section sec_8.10.3 Section sec_8.10.4 8.10 Bursts (transients) on AC and DC mains (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
61<\/td>\n | Section sec_8.11 Section sec_8.11.1 Section sec_8.11.2 Section sec_8.11.3 8.11 Electrostatic discharge (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
62<\/td>\n | Section sec_8.11.4 Section sec_8.12 Section sec_8.12.1 Section sec_8.12.2 Section sec_8.12.3 8.12 Radiated electromagnetic fields (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
64<\/td>\n | Table tab_4 Section sec_8.12.4 Section sec_8.13 Section sec_8.13.1 Section sec_8.13.2 Section sec_8.13.3 8.13 Conducted electromagnetic fields (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
65<\/td>\n | Table tab_5 Section sec_8.13.4 <\/td>\n<\/tr>\n | ||||||
66<\/td>\n | Section sec_8.14 Section sec_8.14.1 Section sec_8.14.2 Section sec_8.14.3 8.14 Surges on signal, data and control lines (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
67<\/td>\n | Section sec_8.14.4 Section sec_8.15 Section sec_8.15.1 Section sec_8.15.2 Section sec_8.15.3 8.15 Surges on AC and DC mains power lines (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, A.5) <\/td>\n<\/tr>\n | ||||||
68<\/td>\n | Section sec_8.15.4 Section sec_8.16 Section sec_8.16.1 Section sec_8.16.2 8.16 Static magnetic field (ISO\u00a04064-1:2014|OIML\u00a0R\u00a049\u20111:2013, 7.2.8) <\/td>\n<\/tr>\n | ||||||
69<\/td>\n | Section sec_8.16.3 Section sec_8.16.4 Section sec_8.16.5 Section sec_8.17 Section sec_8.17.1 Section sec_8.17.2 Section sec_8.17.3 8.17 Absence of flow test <\/td>\n<\/tr>\n | ||||||
70<\/td>\n | Section sec_8.17.4 Section sec_9 Section sec_9.1 Section sec_9.2 Table tab_6 9 Test program for type evaluation 9.1 Number of samples required 9.2 Performance test applicable to all water meters <\/td>\n<\/tr>\n | ||||||
71<\/td>\n | Section sec_9.3 Section sec_9.4 9.3 Performance tests applicable to electronic water meters, mechanical water meters fitted with electronic devices, and their separable parts 9.4 Type evaluation of separable parts of a water meter <\/td>\n<\/tr>\n | ||||||
72<\/td>\n | Section sec_9.5 Section sec_10 Section sec_10.1 Section sec_10.1.1 Section sec_10.1.2 Section sec_10.1.3 9.5 Families of water meters 10 Tests for initial verification 10.1 Initial verification of complete and combined water meters <\/td>\n<\/tr>\n | ||||||
73<\/td>\n | Section sec_10.1.4 Section sec_10.2 Section sec_10.2.1 10.2 Initial verification of separable parts of a water meter <\/td>\n<\/tr>\n | ||||||
74<\/td>\n | Section sec_10.2.2 Section sec_10.2.3 Section sec_10.2.4 Section sec_11 Section sec_11.1 Section sec_11.2 Section sec_11.2.1 11 Presentation of results 11.1 Object of the reports 11.2 Identification and test data to be included in records <\/td>\n<\/tr>\n | ||||||
75<\/td>\n | Section sec_11.2.2 <\/td>\n<\/tr>\n | ||||||
76<\/td>\n | Annex sec_A Annex sec_A.1 Annex sec_A.2 Annex sec_A.3 Annex sec_A.3.1 Annex\u00a0A \n(normative)<\/p>\n Type examination and testing of checking facilities of electronic devices <\/td>\n<\/tr>\n | ||||||
77<\/td>\n | Annex sec_A.3.2 Annex sec_A.3.2.1 Annex sec_A.3.2.2 Annex sec_A.3.2.2.1 <\/td>\n<\/tr>\n | ||||||
78<\/td>\n | Annex sec_A.3.2.2.2 Annex sec_A.3.2.2.3 Annex sec_A.3.3 Annex sec_A.3.3.1 Annex sec_A.3.3.2 Annex sec_A.3.3.2.1 <\/td>\n<\/tr>\n | ||||||
79<\/td>\n | Annex sec_A.3.3.2.2 Annex sec_A.3.4 Annex sec_A.3.4.1 Annex sec_A.3.4.2 <\/td>\n<\/tr>\n | ||||||
80<\/td>\n | Annex sec_A.3.5 Annex sec_A.3.5.1 Annex sec_A.3.5.2 Annex sec_A.3.6 Annex sec_A.3.6.1 <\/td>\n<\/tr>\n | ||||||
81<\/td>\n | Annex sec_A.3.6.2 <\/td>\n<\/tr>\n | ||||||
82<\/td>\n | Annex sec_B Annex sec_B.1 Annex sec_B.2 Annex\u00a0B \n(normative)<\/p>\n Calculating the relative error (of indication) of a water meter <\/td>\n<\/tr>\n | ||||||
83<\/td>\n | Annex sec_B.3 Annex sec_B.3.1 Annex sec_B.3.2 Annex sec_B.3.3 Annex sec_B.3.3.1 <\/td>\n<\/tr>\n | ||||||
84<\/td>\n | Annex sec_B.3.3.2 Annex sec_B.3.3.3 <\/td>\n<\/tr>\n | ||||||
85<\/td>\n | Annex sec_B.3.3.4 Annex sec_B.3.4 Annex sec_B.3.4.1 <\/td>\n<\/tr>\n | ||||||
86<\/td>\n | Annex sec_B.3.4.2 Annex sec_B.3.4.3 <\/td>\n<\/tr>\n | ||||||
87<\/td>\n | Annex sec_B.3.4.4 <\/td>\n<\/tr>\n | ||||||
88<\/td>\n | Annex sec_C Annex\u00a0C \n(normative)<\/p>\n Installation requirements for flow disturbance tests <\/td>\n<\/tr>\n | ||||||
89<\/td>\n | Table tab_v Figure fig_C.1 <\/td>\n<\/tr>\n | ||||||
90<\/td>\n | Annex sec_D Annex sec_D.1 Annex sec_D.2 Annex sec_D.3 Annex\u00a0D \n(normative)<\/p>\n Type evaluation of a family of water meters <\/td>\n<\/tr>\n | ||||||
91<\/td>\n | Figure fig_D.1 <\/td>\n<\/tr>\n | ||||||
92<\/td>\n | Annex sec_E Table tab_w Figure fig_E.1 Annex\u00a0E \n(informative)<\/p>\n Examples of methods and components used for testing concentric water meters <\/td>\n<\/tr>\n | ||||||
93<\/td>\n | Table tab_x Figure fig_E.2 <\/td>\n<\/tr>\n | ||||||
94<\/td>\n | Table tab_y Figure fig_E.3 <\/td>\n<\/tr>\n | ||||||
95<\/td>\n | Annex sec_F Annex sec_F.1 Annex sec_F.2 Annex\u00a0F \n(informative)<\/p>\n Determining the density of water <\/td>\n<\/tr>\n | ||||||
96<\/td>\n | Annex sec_F.3 Table tab_F.1 <\/td>\n<\/tr>\n | ||||||
97<\/td>\n | Annex sec_G Annex sec_G.1 Annex sec_G.2 Annex sec_G.3 Annex\u00a0G \n(informative)<\/p>\n Maximum uncertainties in the measurement of influence factors and disturbances <\/td>\n<\/tr>\n | ||||||
98<\/td>\n | Annex sec_G.4 Annex sec_G.5 Annex sec_G.6 Annex sec_G.7 Annex sec_G.8 <\/td>\n<\/tr>\n | ||||||
99<\/td>\n | Annex sec_G.9 Annex sec_G.10 <\/td>\n<\/tr>\n | ||||||
100<\/td>\n | Annex sec_H Annex sec_H.1 Annex sec_H.2 Annex sec_H.3 Annex\u00a0H \n(informative)<\/p>\n Pressure loss test pressure tappings, hole and slot details <\/td>\n<\/tr>\n | ||||||
101<\/td>\n | Table tab_an Figure fig_H.1 Table tab_ao Figure fig_H.2 <\/td>\n<\/tr>\n | ||||||
102<\/td>\n | Table tab_ap Figure fig_H.3 <\/td>\n<\/tr>\n | ||||||
103<\/td>\n | Annex sec_I Annex sec_I.1 Annex sec_I.2 Table tab_aq Figure fig_I.1 Annex\u00a0I \n(normative)<\/p>\n Flow disturbers <\/td>\n<\/tr>\n | ||||||
104<\/td>\n | Table tab_ar Figure fig_I.2 Table tab_as Figure fig_I.3 <\/td>\n<\/tr>\n | ||||||
105<\/td>\n | Table tab_I.1 Table tab_at Figure fig_I.4 <\/td>\n<\/tr>\n | ||||||
106<\/td>\n | Table tab_I.2 Table tab_au Figure fig_I.5 <\/td>\n<\/tr>\n | ||||||
107<\/td>\n | Table tab_I.3 Figure fig_I.6 Table tab_I.4 <\/td>\n<\/tr>\n | ||||||
108<\/td>\n | Figure fig_I.7 Table tab_I.5 <\/td>\n<\/tr>\n | ||||||
109<\/td>\n | Table tab_av Figure fig_I.8 Table tab_aw Figure fig_I.9 <\/td>\n<\/tr>\n | ||||||
110<\/td>\n | Table tab_ax Figure fig_I.10 <\/td>\n<\/tr>\n | ||||||
111<\/td>\n | Table tab_I.6 <\/td>\n<\/tr>\n | ||||||
112<\/td>\n | Table tab_ay Figure fig_I.11 Table tab_I.7 <\/td>\n<\/tr>\n | ||||||
113<\/td>\n | Figure fig_I.12 Table tab_I.8 <\/td>\n<\/tr>\n | ||||||
114<\/td>\n | Reference ref_1 Reference ref_2 Reference ref_3 Reference ref_4 Reference ref_5 Reference ref_6 Reference ref_7 Reference ref_8 Reference ref_9 Reference ref_10 Reference ref_11 Reference ref_12 Bibliography <\/td>\n<\/tr>\n<\/table>\n","protected":false},"excerpt":{"rendered":" Water meters for cold potable water and hot water – Test methods<\/b><\/p>\n |